Home   Hach   Reagents & Meter

Reagents & Meter ยป DR3900 Laboratory VIS Spectrophotometer with RFID Technology DR3900 Laboratory VIS Spectrophotometer with RFID* Technology DR3900 Laboratory VIS Spectrophotometer and TNTplus nutrients test kits DR3900 Spectrophotometer used with TNTplus test kits in the laboratory DR3900 Laboratory VIS Spectrophotometer with RFID* Technology DR3900 Laboratory Spectrophotometer with flow through application DR3900 Laboratory VIS Spectrophotometer with RFID* Technology DR3900 Laboratory VIS Spectrophotometer with RFID* Technology
Model: DR3900

The DR3900 is a benchtop visible spectrum (320 - 1100 nm), split beam spectrophotometer with over 220 pre-programmed methods optimized for laboratory water analysis. With your daily routine of water analysis in mind, the DR3900 spectrophotometer is optimized for safe processes and accurate results.

The DR3900 is designed to deliver accurate results quickly with Hach's innovative TNTplus® and LCK reagents for fast preparation and easy execution. In fact, it automatically recognizes our Hach reagents, so there is no need to enter method numbers.

Guided Procedures
The DR3900 simplifies water analysis by with step-by-step guides which take users through testing procedures. This helps ensure consistent and accurate results.

RFID Technology
Featuring state-of-the-art Radio Frequency Identification (RFID) technology* allows for hands-free calibration updates, as it automatically recognizes the lot, expiration and calibration curves associated with each Hach chemistry sample. It also allows for traceability for each measurement.

Claros Enabled
This instrument connects to Claros™, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations.

Other Key Features
• Optimized for safe processes and accurate results.
• Large color touchscreen display allows easy setup and navigation.
• RFID technology eliminates sample mix-ups and provides better sample traceability.
• Built with 1 ethernet and 3 USB ports, so easily connects to your computer.
• Over 35 tests available including EPA compliant parameters such as ammonia, COD, chlorine and more.

Specifications

Beam Height 10
Data Logger 2000 measured values (Result, Date, Time, Sample ID, User ID)
Dimensions (H x W x D) 151 mm x 350 mm x 255 mm
Display 7" TFT
Display Resolution WVGA (800 pix x 480 pix)
Display Size 7 inch (17.8 cm)
Display Type Colored touch-screen
Enclosure Rating IP30
Interfaces USB type A (2)
USB type B
Ethernet
RFID module
Light Source Gas-filled Tungsten (visible)
Manual Languages English
French (CDN)
Spanish (SA)
Portuguese (BR)
Chinese
Japanese
Korean
Max. operating humidity 80
Max. Storage Humidity 80
Operating Conditions 10 - 40 °C
Operating Mode Transmittance (%), Absorbance and Concentration, Scanning
Optical System Reference beam, spectral
Photometric Accuracy 5 mAbs @ 0.0 - 0.5 Abs
Photometric Accuracy 2 1 % at 0.50 - 2.0 Abs
Photometric Linearity < 0.5 % - 2 Abs
Photometric Linearity 2 ≤ 0.01 % at >2 Abs with neutral glass at 546 nm
Photometric Measuring Range ± 3.0 Abs (wavelength range 340 - 900 nm)
Power Requirements With external power supply
Power Requirements (Hz) 50/60 Hz
Power Requirements (Voltage) 110 - 240 VAC
Power Supply Benchtop Power Supply
Preprogrammed Methods > 240
Sample Cell Compatibility Rectangular: 10, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch
Scanning Speed > 8 nm/S (in steps of 1 nm)
Source Lamp Gas-filled Tungsten (visible)
Specific Technology RFID for easy method update, sample ID and Certificate of Analysis
Spectral Bandwidth 5 nm
Standard accessories None
Storage Conditions -30 - 60 °C
Stray Light < 0.1 % T at 340 nm with NaNO2
User Interface Languages Bulgarian, Chinese, Croatian, Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Polish, Portuguese - Brasilian, Potuguese, Russian, Serbian, Slovakian, Slowenian, Spanish, Swedish, Turkish
User Programs 100
Warranty 12 months
Wavelength Accuracy ± 1.5 nm (wavelength range 340 - 900 nm)
Wavelength Calibration Automatic
Wavelength Range 320 - 1100 nm
Wavelength Reproducibility ± 0.1
Wavelength Resolution 1
Wavelength Selection Automatic, based on method selection
Weight 4.2 kg
What's included? Includes: Spectrophotometer DR3900, adapter A for 1 in. round and 1 cm square cells, matched pair of 1 in. square glass sample cells, light shield, dust cover, printed basic user manual and benchtop power supply with 115 and 230V power cords.

 

ABOUT US

  • Head Office
  •    3-Syed Mauj Darya Road Lahore
  •    +92-42-37312378/37356445
  •    info@newchemical.com.pk
  • Karachi Office
  •    Humera Center Office # B13,
  •             188/1-A, Block-2, PECHS Karachi.
  •    +92-21-34526689
NEWSLETTER

Subscribe our newsletter gor get notification about new updates, etc.



@2018 NEW CHEMICAL CENTER All RIGHTS RESERVED.